PET/CT QC - Python scripts | Prepared by the EFOMP Working Group (WG) on "PET/CT and PET/MR QC Protocol"
Publication Date: Aug 30th, 2022 21:32, Category: EFOMP Announcements
The EFOMP Working Group on “PET/CT and PET/MR QC Protocol” chaired by Dr Roberta Matheoud (IT) prepared the EFOMP GUIDELINE "Quality control in PET/CT and PET/MR" - software tools user manual.
It is one of the results, where is briefly described general requirements, possible limitations, and etc. (see attachments).
PET/CT QC - Python codes were tested for accuracy with Siemens biograph mCT data. Tests with Philips and GE were done only to a limited extent mainly for compatibility reasons.
07.09.2022 - newer Python version compatibility issues solved by Matteo Bagnalasta (Nuclear Medicine, Foundation IRCCS Istituto Nazionale Tumori, Milan, Italy)!
17.07.2023 - Uniformity analyzer calculation corrected by Matteo Bagnalasta (Nuclear Medicine, Foundation IRCCS Istituto Nazionale Tumori, Milan, Italy) - slice variability checked based on average phantom SUV (not SUV = 1 as in previous version)!
26.07.2024 – RC_analyzer calculation of COV corrected by Evon Smyth (University College Dublin, Dublin, Ireland) is attached!
NEW 24.09.2024 – all scripts corrected. A deprecated Pydicom function read_file was replaced by dcmread function. This prevents any issues with newer Pydicom versions (thanks to Christian Bracco, Mauriziano Hospital, Turin, Italy)
Feedback and issues/bugs reporting are highly appreciated. Please contact Jaroslav Ptacek at jaroslav.ptacek@fnol.cz for this purpose.
Attachments
User Manual Size: 1010,43 KB
WG_PET_CT_QC_Python_24092024 Size: 34,17 KB